|
New AASM Recommendations for Sensors: A Simple Guide for the Sleep
Technologist Prevalence of Sleep Disordered Breathing in Congestive Heart
Failure as Determined by ApneaLink, a Simplified Screening Device
Susan R. Isakson, BS,1 Jennifer Beede, BS,1 Kevin Jiang, BS,1 Nancy
J. Gardetto, MSN, RNP-c,1 Nancy Gordon, MS,3 Eileen Casal, RN, MSN3
and Alan S. Maisel, MD, FACC1,2
Abstract
Background: Sleep
disordered breathing (SDB) is very common in patients with
congestive heart failure (CHF), with some studies reporting an
incidence of 50%. Increased sympathetic activity, caused by
frequent arousals (increased Apnea- Hypopnea Index [AHI]) during
the night has been implicated as a contributing factor in a
four-fold increase in mortality for CHF patients. Many patients are
undiagnosed for SDB due to lack of awareness of the disorder and
access to an easy, reliable screening test. The purpose of this
study is to evaluate the prevalence of SDB in patients with CHF in
the inpatient and outpatient settings using the ApneaLink (AL)
(ResMed Corp, San Diego, CA) screening device.
Methods: 86 patients
with CHF from the VA San Diego Healthcare System were enrolled.
Eligible patients are those with diagnosis of CHF, treatment
naïve for SDB, and symptomatic for sleep apnea (SA) per an SDB
questionnaire created by study personnel. The mean age was 66
years, and the mean BMI was 32 kg/m2. SDB was defined using an AHI
result of =5 events/hour. The prevalence of SA was determined from
the Apnea-Hypopnea Index (AHI) data recorded by the AL. Patients
were included in the analysis if they had a minimum of 4 hours
evaluation time (ET). Determination for the presence of
Cheyne-Stokes respiration (CSR) was performed by evaluation of the
flow signal data.
Results: Seventy percent (61/86) of the patients
met the enrollment criteria. The overall prevalence of SA was 85%
(52/61 subjects). CSR was detected in 33% (20/61 subjects). For the
46 subjects who completed the study in the outpatient setting, the
prevalence of SAand CSR are 83% and 26%, respectively. In the
inpatient setting, 15 patients with an admission diagnosis of CHF
completed the study with a prevalence of 93% and 53% for SA and
CSR, respectively.
Conclusions: SDB appears to be relatively common
in the CHF population, regardless of age. Preliminary results
provide supportive evidence for the clinical need to identify,
diagnose and treat the CHF patient with SDB. The AL is a relatively
simple test to administer to patients and easy to use by patients
in the home setting. A larger sample size of hospital patients
using the device is required to evaluate utility in the inpatient
setting.
-------------------------
1. San Diego VA Medical Center, San Diego, CA.
2. University
of California, San Diego, CA.
3. ResMed Corp, San Diego, CA.
Introduction
Sleep disordered breathing (SDB) is a common
underlying disease in patients with congestive heart failure (CHF)
and left ventricular dysfunction, with some studies reporting a
prevalence from 50% to as high as 70%.1,2 However, SDB is
frequently overlooked in the management of heart failure patients.
SDB incorporates both central sleep apnea (CSA, including Cheyne-Stokes respiration or periodic breathing) and obstructive
forms of sleep apnea. Cheyne-Stokes respiration (CSR) is described
as recurring episodes of crescendo hyperventilation and decrescendo
hypoventilation with periodic apnea (cessation of breathing) and
hypopnea (reduction of airflow) events that can lead to excessive
arousals, oxygen desaturation, and changes in intrathoracic
pressure.3–5 This can result in adverse cardiac effects such
as ischemia and arrhythmias from sympathetic nervous activation,
increased blood pressure, and increased myocardial oxygen
demand.3–13 The National Institutes of Health (NIH) stated in
their JNC-7 High Blood Pressure Guidelines that sleep apnea was an
identifiable cause of hypertension. Several studies have shown that
patients with SDB have an increased morbidity and worsened
prognosis.14–16 Furthermore, the presence of SDB has been
correlated with an increased risk of developing CHF.17
Sleep apnea,
primarily CSA and CSR, is known to be prevalent in patients with
CHF. Schulz et al. found that 71% of patients with CHF had an AHI
> 10 events/hour.18 However, many patients in this population
remain undiagnosed because of several obstacles. First, because
symptoms of SDB often mimic those of heart failure, patients with
these symptoms may be thought to just be suffering from worsened
heart failure, and as such, are not referred for sleep testing.
Second, because traditional sleep testing is time consuming and
cumbersome, many practitioners and patients may be reluctant to
take this step.
Early diagnosis of SDB and initiation of therapy
may significantly improve the prognoses and reduction of symptoms
in patients with heart failure. A study of nocturnal home oxygen
therapy for treatment of sleep apnea showed concurrent improvements
in AHI and LVEF.19 CPAP therapy has also been shown to have a
beneficial effect on LVEF in patients with SA. In a study by Egea
et al., mean LVEF in patients treated with CPAP increased from
28.0% to 30.5% (p<0.001) after therapy, whereas patients in the
sham-CPAP group did not show any improvement in LVEF.20 In a
similar study, Kaneko et al. examined the effects of CPAP on LVEF,
blood pressure and heart rate in patients with dilated
cardiomyopathy. They found that the CPAP group had significant
increases in LVEF after one month of CPAP, as well as a mean
decrease in systolic blood pressure of 10 mmHg (p=0.02) and
decrease in heart rate of 4 beats per minute (p=0.007). The control
group did not show improvement in any of these variables.21
Additionally, for patients with CHF and CSA/CSR, studies have shown
that Adaptive Servo Ventilation (VPAP Adapt SV™, ResMed Inc,
Poway, CA) improves not only AHI, SpO2, and sleep architecture, but
also may improve VO2 max, six minute walk, quality of life, LVEF
and other important cardiovascular outcomes for heart failure
patients.22–26
With recent evidence implicating a positive
impact on CHF symptoms with treatment of SDB, there may be an
increasing need to screen for SDB.18 Primary care providers as well
as Cardiologists have been frustrated by the lack of a
simple-to-use, yet sensitive and specific device to detect SDB in
their patients. The major aim of the present study was to determine
if assessing for SDB was feasible in a population of stable
outpatient CHF patients using a portable apnea detector, the
ApneaLink™ (AL) device. The secondary aims of the study were
to determine prevalence of SDB in this population, clinician and
patient ease of use, as well as the cost-effectiveness of the
portable device in a real-world setting.
Methods
Patient Enrollment
From September 2005 through January 2007, 86 patients with a known
diagnosis of left ventricular dysfunction with CHF and who were
being followed in the outpatient specialized cardiology clinic or
being treated for decompensation in the hospital at the Veterans
Affairs Medical Center, La Jolla, California, were enrolled. The
inclusion criteria were age >18 years, diagnosis of CHF,
symptomatic for possible SDB (as measured by using a questionnaire
created for the study which asked about specific symptoms of SDB
including snoring, daytime sleepiness, waking up at night feeling
short of breath and others, or by clinical suspicion of an
examining physician), treatment naïve for SDB, and ability to
provide informed consent. The study was approved by an
Institutional Review Board. Patients were excluded from the study
if they were asymptomatic for SDB, had a history of treatment for
SDB or were currently receiving positive airway pressure therapy.
Other exclusion criteria were outpatients requiring home oxygen
therapy, or inpatients requiring >4 L/minute of oxygen or
receiving oxygen via face mask.
Of the 120 patients screened for
this study, 86 were eligible for enrollment. Of the 86 patients
consented, approximately 75% were outpatients and 25% inpatients.
Seventy one of these 86 subjects completed the study and 15
withdrew, stating that the primary reason for withdrawal was lack
of interest in continuing their participation. Sixty one patients
met the criteria of 4 or more hours of study time and are included
in the analyses. See Figure 1.
ApneaLink Screening Device
The ApneaLink™ (ResMed Inc, Poway, CA) is a portable,
battery-powered, respiratory pressure sensor-based sleep apnea
testing system and provides detailed recordings of respiratory flow
during sleep (Figure 2). Patients were instructed on the proper use
of the device and nasal cannula at the time of consent and were provided with an
instruction sheet that included both diagrams and written instructions. The
recorder was fastened with a belt onto the patient's chest.
All relevant
respiratory information, including breaths per minute, frequency of
apneas and hypopneas and snoring events, during sleep was collected
via nasal pressure cannula. Study personnel suggested fitting the
ApneaLink device 30 minutes prior to going to sleep to get
comfortable with the sensation of the cannula, and turning the
ApneaLink on just before going to sleep. Patients were asked to
wear the ApneaLink device overnight and return the device to study
personnel for download of the data. Using ApneaLink review software
installed on a computer, study personnel downloaded stored data
from the ApneaLink device and generated a report which included
apnea-hypopnea index (AHI) scores. If the evaluation period was
less than 4 hours, patients were contacted and asked to repeat the
study. An example of a standard report is illustrated in Figure 3.

Statistical Analyses
The prevalence of SA was determined from the ApneaLink results at AHI levels of =5 events/hour, =10, =15, and
5–14 (dependent on qualifying co-morbidities) with the use of
confidence interval testing. The analyses included patients who had
an evaluation period of 4 or more hours of study time on the
ApneaLink device. The ApneaLink report was evaluated for the
presence of Cheyne-Stokes respiration (CSR) and the prevalence of
CSR. Prevalence rates along with 95% confidence intervals are
reported. The analyses presents the cohort combined, as well as
split out by inpatients/outpatients.
In addition, ease-of-use of
the ApneaLink device and acceptability were evaluated using a
5-point Likert scale with '5' representing Excellent and '1' representing Poor.
Time and cost analyses related to use of the device were
completed by an outside economic and health outcomes consultant.
The variables were assessed by evaluating the set-up time and
report generation features of the ApneaLink device.
Results
Of the
86 patients who consented to participate in the study, 61 (46
outpatients and 15 inpatients) had 4 or more hours of recorded
evaluable data for assessment of the prevalence of SA. Demographic
information is presented in Table 1. The mean age of the 61
evaluable patients was 65 years (range 38–84) and the mean
BMI was 32.0 kg/m2 (range 21.4–51.1). 98% of the cohort were
male and 70% were white.
Medical history information is presented
in Table 2. Most patients were classified as having moderate CHF,
with 46% (28/61) of patients in NYHA Class II and 33% (20/61) in
Class III. Diagnosis of CHF was made an average of 7.1 years prior
to enrollment in this study (range <1–26 years). A history
of systolic heart failure was present in 95%, (52/61) of the
patients, and 80% (49/61) had a history of hypertension. The most
common etiology of CHF was ischemia (56%, 34/61). With respect to
sleep history, the most frequently reported sleep characteristics
were excessive sleepiness (74%), feeling un-refreshed after
sleeping (74%), and frequent snoring (62%).
Prevalence rates of SA
based on AHI along with 95% confidence intervals are presented in
Table 3. Prevalence of obstructive sleep apnea was 85% (SA defined
as an AHI of 5 events/hour or greater as recorded by the eaLink). Prevalence in inpatients was found to be slightly
higher (93%) than for outpatients (83%). Forty-nine percent of
patients had an AHI of 15 events/hour or greater, and 36% had an
AHI between 5–14 events/hour with qualifying co-morbidities:
documented symptoms of excessive daytime sleepiness, impaired
cognition, mood disorders or insomnia, or documented hypertension,
ischemic heart disease, or history of stroke, per CMS criteria.
This sample size (N=61) was able to estimate the assumed SDB rate
of 50% with a precision of 12.5%. No significant difference was
seen in prevalence of SDB between obese and non-obese patients.
The
prevalence of Cheyne-Stokes Respiration (CSR) is presented in Table 4.
Thirty-three percent (20/61) of patients were diagnosed with CSR, with
inpatients having a much higher prevalence than outpatients (53% vs. 26%,
p=0.051) 80% of subjects with CSR had an AHI of 15 events/hour or greater
(Table 5).
One of the objectives of the study was to perform a
time/cost analysis related to the use of the device. Overall, the
user (study coordinator) required 3.1 minutes to set up the ApneaLink device and 2.1 minutes to generate the report. A simple
model was constructed to compare the labor cost for performing the
ApneaLink with the labor cost for performing a similar test - an
ECG. The model applies estimated times to perform each component of
the two procedures to national hourly wage and benefit rates.
Hourly rates ranged from a low of $10.00/hour (nurse's aide) to a
high of $25.96/hour (registered nurse, US Bureau of Labor
Statistics (BLS) National Compensation Survey), while the average
fringe benefit rate was 28%. Based on model results, the labor cost
is roughly equivalent for the two procedures, with the ApneaLink
costing slightly less. For example, using the national hourly wage
and including fringe benefits for a RN ($33.23) the cost to perform
an ApneaLink test is $2.77 (5 minutes to perform complete test) vs.
$3.88 (7 minutes to perform an ECG test).
A further
objective was to determine the ease of use and acceptability of ApneaLink as a screening tool in CHF patients using the 5-point
Likert scale mentioned in the methods section. Both patients and
clinicians showed a very high rate of acceptance, with overall
patient acceptance rated above average or excellent by 95% of the
subjects and overall performance rated above average or excellent
by 100% of respondents.
For questions relating to the clinician's
experience were rated above average or excellent by 100% of
respondents. For questions relating to the patient's experience
(unobtrusiveness and ease of use), most items were rated above
average or excellent by more than 90% of respondents, with the
lowest response at 87%. Little difference was found between
inpatients and outpatients in the rate of positive responses. All
criteria were rated above average or excellent by at least 83% of
inpatients, and by at least 87% of outpatients. In the category
Patient Ease of Use, outpatients were especially positive in their
ratings: 100% of outpatients and 86% of inpatients rated usefulness
of supplied patient instructions as above average or excellent;
operations of controls and buttons was rated above average or
excellent by 96% of outpatients and 85% of inpatients. The higher
level of positive ratings may reflect a greater reliance on
instructions and controls in the outpatient group.
Discussion
The
results of this study indicate that when CHF patients in a Veterans
population are prescreened for symptoms that might represent SDB,
the prevalence of SAis high (85%), with a slightly higher
prevalence in inpatients versus outpatients. CSR was present in
approximately one third of the patients. There was a high level of
satisfaction expressed by both patients and clinicians in terms of
ease of use of the ApneaLink. Additionally, the cost analysis
proved ApneaLink to be cost-effective for screening use in the
inpatient and outpatient settings.
Our results bring into focus the
degree to which SBD may exist in a Veteran CHF population. The high
prevalence of SDB in this population and the physiological insults
experienced by patients with SDB (ischemia and arrhythmias from
sympathetic nervous activation, increased blood pressure, and
increased myocardial oxygen demand) suggest that on both an
individual and epidemiological level, SDB may significantly impact
the progression of CHF and the prognosis of these
patients.3–5
As such, the benefits of a convenient and
readily available sleep-screening tool are several-fold. The device
used in this study, the ApneaLink, has been shown to have strong
clinical utility for providers to screen their patients suspected
of having SDB. The accuracy of the ApneaLink was validated in a
study comparing ApneaLink recordings with simultaneously collected
data from polysomnography recordings. The ApneaLink was shown to be
highly sensitive and specific (91% and 95%) when AHI data from the
two measurements were compared. Additionally, ApneaLink results
from screening in patients homes showed high sensitivity and
specificity when compared with the polysomnography laboratory
results.27
The results from ApneaLink testing can be used to
stratify patients according to need for further testing in a
polysomnography lab in the case of SDB with CSR, with subsequent
Adaptive Servo-Ventilation (ADV) therapy and also to identify
patients who would benefit from immediate titration and/or
initiation of CPAP therapy for SDB without CSR. Inpatients being
treated for decompensated heart failure that undergo ApneaLink
testing during the hospital stay could potentially be started on
CPAP, or ASV depending on the level of CSR detected, as an adjunct
to their CHF management.
Further studies examining the clinical
utility of ApneaLink should encourage more widespread screening in
different populations. The new model of the ApneaLink device
includes a pulse oximeter, and studies testing this device will be
helpful to determine how the presence of pulse rate and pulse
oximetry data improve the quality of assessments. Additional
studies comparing results from ApneaLink and polysomnography are
likely to be helpful in determining the accuracy of the ApneaLink
in measuring severity of SDB.
Study Limitations
The number of
patients enrolled in the study was relatively small due to a
limited patient population from which to recruit, as well as a
small number of patients admitted to the hospital who fit the
inclusion/exclusion criteria. Additionally, seven patients were
included in the study that did not answer yes to at least two
screening questions. The investigator made the decisions to enroll
these patients based on clinical suspicion of possible SDB and the
potential benefit of sleep screening to the patient.
Inability to
obtain a full 4-hour recording was a common problem encountered
during the study. The reasons most often cited for less than 4
hours of evaluation time were difficulty with the on/off switch and
some difficulty keeping the nasal cannula in place. This feedback
was passed on to the manufacturer and was added to improvements in
their ApneaLink with Oximetry product that is now released to the
market but was unavailable during this trial. The ApneaLink with
Oximetry meets the CMS definitions of a Type IV Home Diagnostic
Device with the following measurement signals: respiratory flow,
heart rate and oximetry. Several subjects withdrew from the study
because they were not interested in repeating the screening.
Reasons cited by these patients included lack of a return
appointment at VA hospital, trouble sleeping through the night, and
no recording after two attempts.
Conclusion
The ApneaLink is an
easy-to-use, portable, cost-effective way to assess SDB in patients
with heart failure. It provides a useful way to differentiate SDB
into a group with CSR and a group without CSR, which has important
treatment implications for the heart failure patient population.
The high prevalence of SDB in this population suggests that heart
failure patients should be routinely screened for SDB.
References
1. Javaheri, S., et al., Sleep apnea in 81 ambulatory male patients
with stable heart failure. Types and their prevalences,
consequences, and presentations. Circulation, 1998. 97(21): p.
2154–2159.
2. Ferrier, K., et al., Sleep-disordered breathing
occurs frequently in stable outpatients with congestive heart
failure. Chest, 2005. 128(4): p. 2116–2122.
3. Bradley, T.D.
and J.S. Floras, Sleep apnea and heart failure: Part I: obstructive
sleep apnea. Circulation, 2003. 107(12): p. 1671–1678.
4. Bradley, T.D. and J.S. Floras, Sleep apnea and heart failure: Part
II: central sleep apnea. Circulation, 2003. 107(13): p. 1822–1826.
5. Leung, R.S., et al., Provocation
of ventricular ectopy by cheynestokes respiration in patients with
heart failure. Sleep, 2004. 27(7): p. 1337–1343.
6. Andreas,
S., et al., Cheyne-Stokes respiration and prognosis in congestive
heart failure. Am J Cardiol, 1996. 78(11): p. 1260–1264.
7. Hanly, P.J. and N.S. Zuberi-Khokhar, Increased mortality associated
with Cheyne-Stokes respiration in patients with congestive heart
failure. Am J Respir Crit Care Med, 1996. 153(1): p. 272–276.
8. Sin, D.D., et al., Risk factors for central and obstructive
sleep apnea in 450 men and women with congestive heart failure. Am
J Respir Crit Care Med, 1999. 160(4): p. 1101–1106.
9.
Bradley, T.D. and J.S. Floras, Pathophysiologic and therapeutic
implications of sleep apnea in congestive heart failure. J Card
Fail, 1996. 2(3): p. 223–240.
10. Javaheri, S., Central sleep
apnea-hypopnea syndrome in heart failure: prevalence, impact, and
treatment. Sleep, 1996. 19(10 Suppl): p. S229–S231.
11. Yamashiro, Y. and M.H. Kryger, Review: sleep in heart failure.
Sleep, 1993. 16(6): p. 513–523.
12. Franklin, K.A., et al.,
Sleep apnea and nocturnal angina. Lancet, 1995. 345(8957): p.
1085–1087.
13. Javaheri, S., Sleep disorders in systolic
heart failure: a prospective study of 100 male patients. The final
report. Int J Cardiol, 2006. 106(1): p. 21–28.
14. Ancoli-Israel, S., et al., Comparison of patients with central
sleep apnea. With and without Cheyne-Stokes respiration.Chest,
1994.106(3): p. 780–786.
15. Javaheri, S., et al., Central
sleep apnea, right ventricular dysfunction, and low diastolic blood
pressure are predictors of mortality in systolic heart failure. J
Am Coll Cardiol, 2007. 49(20): p. 2028–2034.
16. Wang, H., et
al., Influence of obstructive sleep apnea on mortality in patients
with heart failure. J Am Coll Cardiol, 2007. 49(15): p.
1625–1631. 17. Shahar, E., et al., Sleep-disordered breathing
and cardiovascular disease: cross-sectional results of the Sleep
Heart Health Study. Am J Respir Crit Care Med, 2001. 163(1): p.
19–25.
18. Schulz, R., et al., Sleep apnea in heart failure. Eur Respir J, 2007. 29(6): p. 1201–1205.
19. Seino, Y., et
al., Clinical efficacy and cost-benefit analysis of nocturnal home
oxygen therapy in patients with central sleep apnea caused by
chronic heart failure. Circ J, 2007. 71(11): p. 1738–1743.
20. Egea, C.J., et al., Cardiac function after CPAP therapy in
patients with chronic heart failure and sleep apnea: A multicenter
study. Sleep Med, 2007.
21. Kaneko, Y., et al., Cardiovascular
effects of continuous positive airway pressure in patients with
heart failure and obstructive sleep apnea. N Engl J Med, 2003.
348(13): p. 1233–1241.
22. Zhang, X.L., et al., Efficacy of
adaptive servoventilation in patients with congestive heart failure
and Cheyne-Stokes respiration. Chin Med J (Engl), 2006. 119(8): p.
622–627.
23. Philippe, C., et al., Compliance with and
effectiveness of adaptive servoventilation versus continuous
positive airway pressure in the treatment of Cheyne-Stokes
respiration in heart failure over a six month period. Heart, 2006.
92(3): p. 337–342.
24. Morgenthaler, T.I., et al., Adaptive
servoventilation versus noninvasive positive pressure ventilation
for central, mixed, and complex sleep apnea syndromes. Sleep, 2007.
30(4): p. 468–475.
25. Pepperell, J.C., et al., A randomized
controlled trial of adaptive ventilation for Cheyne-Stokes
breathing in heart failure. Am J Respir Crit Care Med, 2003.
168(9): p. 1109–1114.
26. Teschler, H., et al., Adaptive
pressure support servo-ventilation: a novel treatment for
Cheyne-Stokes respiration in heart failure. Am J Respir Crit Care
Med, 2001. 164(4): p. 614–619.
27. Erman, M.K., et al.,
Validation of the ApneaLink for the screening of sleep apnea: a
novel and simple single-channel recording device. J Clin Sleep Med,
2007. 3(4): p. 387–392.
Journal
List
|